Detecting Preload Changes Using Seismocardiography

Ahmad Agam1, Peter Søgaard2, Charlotte Burup Kristensen3, Rasmus Møgelvang4, Samuel Schmidt5
1Aalborg University and Aalborg University Hospital, 2Aalborg University Hospital, 3Copenhagen University Hospital Rigshospitalet, Lund University Hospital, 4Copenhagen University Hospital, 5Aalborg University


Abstract

Abstract

Introduction: Seismocardiography (SCG) is a non-invasive tool that utilizes the chest wall vibrations from the beating heart to measure cardiac measures using a highly sensitive accelerometer. SCG offers continuous measurement of cardiac function that could be utilized for monitoring, diagnostic assessments, and prognostic health checks. Aim: The aim of this study was to investigate if the SCG could detect an increase in preload by fast saline infusion.

Methods: Twenty-six subjects were included in this study, sixteen with cardiac diseases such as hypertrophic cardiomyopathy, dilated cardiomyopathy, aortic valve disease or ischemic heart disease (age 45.8±17.7 years and 93% male) and ten subjects without known cardiac conditions (age 42.1±14.4 years and 70% male). Ensemble average SCG heartbeat signals were constructed and utilizing a fiducial point detection algorithm, the amplitudes and time intervals of the SCG fiducial points were identified. Wilcoxon signed rank test was used to test for significant difference pre and post infusion.

Results: Post infusion SCG was not obtained in two subjects, therefore only twenty-four subjects were included post infusion and twenty-six pre infusion. Post infusion the diastolic SCG amplitudes Dd increased 24% (p=.016), Ed increased 47% (p = <.000) and the systolic Ls increased 20% (p = <.015). Diastolic time intervals Bd to Fd decreased with 9% (p = .010), and Bd to Ed decreased 4% (p = .040). The same was observed for early systolic time intervals Es to Gs that decreased 3% and Es to Ls that decreased 8% (p = <.000) However, the mid systolic interval Gs to Bd increased 5% and Ls to Bd increased 12%

Conclusion: The results found in this study demonstrates a significant difference between the SCG measures pre fluid and post fluid infusion, thus indicating that the SCG could potentially be utilized to detect preload changes.